NSN 4110-01-497-3082
Part Details | BLOOD PLASMA MECHANICAL FREEZER
4110-01-497-3082
Alternate Parts: 020491, 02-0491, 4110NCM010174, UFP430A, 4110-01-497-3082, 01-497-3082, 4110014973082, 014973082
Supply Group (FSG) | NSN Assigned | NIIN | Item Name Code (INC) |
---|---|---|---|
41 | MAY 16, 2002 | 01-497-3082 | 38484 ( FREEZER, MECHANICAL, BLOOD PLASMA ) |
REFERENCE DRAWINGS & PICTURES
Cross Reference | NSN 4110-01-497-3082
Part Number | Cage Code | Manufacturer |
---|---|---|
02-0491 | 89875 | DEFENSE LOGISTICS AGENCYDIRECTORATE OF MEDICAL MATERIEL |
4110NCM010174 | 64616 | DEFENSE MEDICAL STANDARDIZATIONBOARD |
UFP430A | 1FKX5 | REVCO TECHNOLOGIES INC |
Technical Data | NSN 4110-01-497-3082
Characteristic | Specifications |
---|---|
OUTSIDE SHELL MATERIAL | STEEL, CORROSION RESISTING |
COLOR | GRAY |
CAPACITY AND LOCATION | AXA4.900 CUBIC FEET NOMINAL |
OVERALL WIDTH | 24.000 INCHES NOMINAL |
OVERALL LENGTH | 26.000 INCHES NOMINAL |
OVERALL HEIGHT | 33.375 INCHES NOMINAL |
DOOR QUANTITY | 1 |
OPERATION METHOD | COMPRESSION |
CURRENT TYPE | DC |
VOLTAGE RATING | 115.0 VOLTS NOMINAL |
FREQUENCY RATING | 60.0 HERTZ NOMINAL |
EVAPORATOR | NOT INCLUDED |
USAGE DESIGN | COMMERCIAL |
SPECIAL FEATURES | PRE-SET -30 DEGREES C TEMPERATURE SETPOINT; AUDIBLE/VISUAL WARNING FOR OVER TEMPERATURE, UNDERTEMPERATURE AND POWER FAILURE, WITH VISUAL STATUS REPORTS ON CRITICAL FUNCTIONS; A BATTERY BACK-UP ASSURES SYSTEM READINESS; ICON-BASED CONTROLS AND DISPLAYS INCLUDE GRAPHIC REFERENCES FOR UNIVERSAL OPERATIONS; PUSH BUTTON ALARM TEST VERIFIES SENSORS CONTINUITY; SETPOINT SECURITY WITH KEY OPERATED TRIPLE-POSITON MASTER SWITCH CONTROLS POWER TO ALL SYSTEMS FOR ADDED SAFETY |