NSN 4110-01-424-9408
Part Details | BLOOD PLASMA MECHANICAL FREEZER
4110-01-424-9408
Alternate Parts: 960136, 96-0136, 4110-01-424-9408, 01-424-9408, 4110014249408, 014249408
Supply Group (FSG) | NSN Assigned | NIIN | Item Name Code (INC) |
---|---|---|---|
41 | FEB 08, 1996 | 01-424-9408 | 38484 ( FREEZER, MECHANICAL, BLOOD PLASMA ) |
REFERENCE DRAWINGS & PICTURES
Cross Reference | NSN 4110-01-424-9408
Part Number | Cage Code | Manufacturer |
---|---|---|
96-0136 | 89875 | DEFENSE LOGISTICS AGENCYDIRECTORATE OF MEDICAL MATERIEL |
Technical Data | NSN 4110-01-424-9408
Characteristic | Specifications |
---|---|
OUTSIDE SHELL MATERIAL | STEEL, CORROSION RESISTING |
OVERALL WIDTH | 36.000 INCHES NOMINAL |
OVERALL LENGTH | 24.000 INCHES NOMINAL |
OVERALL HEIGHT | 25.000 INCHES NOMINAL |
DOOR QUANTITY | 1 |
CURRENT TYPE | AC |
VOLTAGE RATING | 120.0 VOLTS NOMINAL |
FREQUENCY RATING | 60.0 HERTZ NOMINAL |
SPECIAL FEATURES | UNDERCOUNTER DESIGN. 5.4 CU. FT. MINUS 30 DEGREE TEMP. RATING. STORAGE CAPACITY: A MIN. OF 75 FILLED PLASMA BAGS. THREE (3) STAINLESS STEEL (CRES) ROLL-OUT DRAWERS. THE TOTAL EXTERIOR CUBE SHALL NOT EXCEED 12.5 CUBIC FT. SHALL HAVE AN ALARM SYSTEM WITH AND AUDIBLE AND VISUAL SIGNAL FOR VIOLATION OF TEMP. REQUIREMENTS AND LOSS OF ELECT. POWER. THE ALARM SYSTEM SHALL HAVE A RECHARGEABLE BATTERY BACKUP. COMES WITH A MIN. OF ONE (1) YEAR SUPPLY OF TEMPERATURE RECORDING CHARTS |
CUBIC MEASURE | 12.5000 CUBIC FEET |
SUPPLEMENTARY FEATURES | THE FREEZER SHALL BE EQUIPPED WITH A CONTINUOUSLY RECORDING THERMOMETER CAPABLE OF RECORDING TEMP. REQUIREMENTS FOR SEVEN DAYS ON ONE CHART. |